Below are the release notes for NOTOCORD-hem version 4.3 containing all changes since NOTOCORD-hem version 4.2.

TypeDefinition
NNew functionality
RRemoved functionality
CCorrection of a defect
KKnown issue

  

fromto

Oops, it seems that you need to place a table or a macro generating a table within the Table Filter macro.

The table is being loaded. Please wait for a bit ...

TypeTopicVersionDescriptionIDRevised on
NVSH10a/bVSH10 PatchCompatibility with Hikvision DS-4308HCVI-E video acquisition card.HEM-2602017/09/18
KEXP10vVSH10 PatchExporting the video signal from a vertical zone in a CTD60v is not possible when the audio has been recorded with the Hikvision DS-4004HCI and DS-4008HCI cards (old video acquisition cards).HEM-6192017/09/18
NDSI71a/b/c/d/e4.3.0.75Compatibility with DSI Hardware Configuration 1.8 (DSI-Link 1.8.16235.1).HEM-2182017/03/17
NDSI71a/b/c/d/e4.3.0.75The sampling rate is indicated for each implant channel in the server properties window.HEM-3032017/03/17
NDSI71a/b/c/d/e4.3.0.75Audit trail and configuration report include implant channel configuration.HEM-2982017/03/17
KDSI71a/b/c/d/e4.3.0.75Calibration values for implants connected to MX2 devices are not available in the Audit trail and Configuration report (DSI-Link 1.8 limitation).HEM-3352017/03/17
KDSI71a/b/c/d/e4.3.0.75Filter Cutoff activation/deactivation for PhysioTel Digital implants is not available in the Audit trail and Configuration report (DSI-Link 1.8 limitation).HEM-3932017/03/17
NDSI71a/b/c/d/e4.3.0.75A dedicated DSI-API Log file can be activated in the server Configuration window in Diagnostics menu.HEM-1702017/03/17
NDSI71a/b/c/d/e4.3.0.75The channel name of implants connected to a MX2 is constituted by the channel number and the Signal type, e.g. "Ch 1 LVPressure", "Ch 2 ECG".HEM-1292017/03/17
CDSI71a/b/c/d/e4.3.0.75OK button on the server properties window is accessible when the server has been closed a first time without configuring any implant.HEM-2112017/03/17
CDSI71a/b/c/d/e4.3.0.75Changes of the Default ambient pressure is now correctly reported in the Audited trail.HEM-3332017/03/17
CDSI71a/b/c/d/e4.3.0.75The Offset feature is no longer displaying an error when setting a value and is working properly.HEM-762017/03/17
KDSI71a/b/c/d/e4.3.0.75DSI Event are not always displayed during acquisition without modifying manually the display scaling or zoom level.HEM-3282017/03/17
RDSI71a/b/c/d/e4.3.0.75The default ambient pressure value displayed in the Pressure correction column for each pressure channel has been removed.HEM-3292017/03/17
CDSI71a/b/c/d/e4.3.0.75Filtering options for APR device signal has been disabled. Raw APR is always used to correct implant pressure signals.HEM-4042017/03/17
NSTE10a/b/c/d4.3.0.75Stellar Telemetry acquisition server - Prototype version.STE102017/03/17
NSTE30a/b/c/d4.3.0.75Stellar Telemetry acquisition server - Prototype version.STE302017/03/17
NVSH10a/b4.3.0.75The Bitrate of the exported video signal has been increased to improve the quality.HEM-492017/03/17
KMON10a4.3.0.75Video signals acquired using NOTOCORD-hem 4.2 on Windows XP cannot be monitored on a 4.3 version under Windows 7.HEM-282017/03/17
CMON10a4.3.0.75Correction of an error that was precluding video channels selection on NOTOCORD-VideoMonitoring.HEM-2322017/03/17
CVSB10a4.3.0.75The warning stating that VSB10a module is not functional no long appears when opening a file containing VSB10a module on a review system.HEM-512017/03/17
CECG51a4.3.0.75Correction of 3 specific cases inducing an ECG51 "stream read error". The stream is corrected at file opening.HEM-782017/03/17
CDTS604.3.0.75All labels are in English when the computer regional setting is English.HEM-522017/03/17
NAMG50a4.3.0.75It's indicated in AccessManager that a local IT administrator have access to configure AccessManager.HEM-1742017/03/17
CAMG50a4.3.0.75Audit Trail Report and Audit Trail Extraction in Excel include the old and new file paths when a file is opened out of the workfile directory.HEM-2422017/03/17
CAMG50a4.3.0.75The error message when NOTOCORD Site Manager (NSM Local) service did not start or is not present has been modified to display the correct information.HEM-1302017/03/17
CSTA10a4.3.0.75The calculation on analysis period delimited by 2 successive Starting triggers when no Ending trigger is connected is corrected.HEM-262017/03/17
NInstaller4.3.0.75Microsoft .NET 4.6.2 Full Profile is a prerequisite to install NOTOCORD-hem™ and replaces Microsoft .NET 4.5 Full Profile in the prerequisites folder.HEM-3722017/03/17
CMain window4.3.0.75NOTOCORD-hem™ Main window is now maximised and in the foreground when this layout is saved as default configuration.HEM-502017/03/17
NRME10s4.3.0.74New reference mark editor for seizure - Prototype version.HEM-1062016/05/31
CDSI71a/b/c/d/e4.3.0.74DSI71 Pressure correction parameter is not reset for unmodified channels when closing DSI Hardware Configuration windows.HEM-1022016/05/31
CDSI71a/b/c/d/e4.3.0.74A number is added at the end of each channel when DSI API returns the same name to avoid channels name duplication.HEM-982016/05/31
CDSI71a/b/c/d/e4.3.0.74APR options has been removed from Diagnostics menus.HEM-1032016/05/31
CDSI71a/b/c/d/e4.3.0.74DSI71 server is no longer crashing at file closure when it fails to connect to an IPC port.HEM-272016/05/31
CSTE10a/b/c/d4.3.0.74
Incomplete data packets (< 100 elements) are removed from the signal recordings. 
HEM-1112016/05/31
RLVP30a4.3.0.74LVP30a module is replaced by LVP31a. LVP30a module is now in Legacy tab.HEM-972016/05/31
RDTT10a4.3.0.74DTT10a module is replaced by DTV10a. DTT10a module is now in Legacy tab.HEM-972016/05/31
RPLY30a4.3.0.74PLY30a module is replaced by PLY30m. PLY30a module is now in Legacy tab.HEM-972016/05/31
NDocumentation4.3.0.73New link "Online documentation" replaces User manual, Reference documentation, Release Notes, and My NOTOCORD links in NOTOCORD-hem help menu.  HEM142016/04/06
NDocumentation4.3.0.73Online documentation website http://confluence.notocord.com now replaces my.notocord.com website.HEM142016/04/06
NDSI71a/b/c/d/e4.3.0.73The help menu and the help buttons in DSI71 properties window now open DSI PhysioTel Digital help file.HEM192016/04/06
NLVP32a4.3.0.73New Left ventricular pressure analyzer that detects ESP point and calculates dP/dt at a user-defined pressure, Isovolumic contraction and relaxation times, Diastolic filling time, Ejection pressure and Diastolic filling pressure.LVP32a2016/04/06
CKRN434.3.0.73The acquisition starts at the correct date and time when the workstation has not been restarted since 25 days.HEM202016/04/06
CDSI60a/b/c/d4.3.0.73Support for DSI OpenART 4.33 and 4.34 on 32-bit systems and OpenART 4.34 on 64-bit systems.HEM232016/04/06
CSTE10a/b/c/d4.3.0.73The implant memory management has been corrected to avoid acquisition interruption.HEM222016/04/06
CLVP31a4.3.0.73With this correction, the module now analyzes the whole pressure signal when Levenberg-Marquardt method is enabled on particular LVP signals.WI55382016/04/06
CLVP31a4.3.0.73r2 calculation with Weiss, N.Z.A., Levenberg-Marquardt and Logistic methods has been corrected.WI55262016/04/06
CLVP31a4.3.0.73dP/dt Max and dP/dt Min detection have been corrected and are now independent from the signal offset.HEM102016/04/06
CLVP31a4.3.0.73The Mean pressure calculation has been corrected by excluding the next EDP point from the calculation.HEM162016/04/06
RInstaller4.3.0.73User manual, Reference documentation and Release Notes are no longer installed locally.HEM142016/04/06
RDSI71a/b/c/d/e4.3.0.73Demo1, Demo2 and Demo3 options have been removed from Configure sources and Diagnostics menus.HEM182016/04/06
NSTE10a/b/c/d4.3.0.72New Acquisition server for Stellar telemetry systems with GLP features and up to 64 implants.STE102015/11/03
CMarkers4.3.0.72Number of a marker changed when the X axis is rescaled. The issue is fixed.WI53892015/11/03
KUser Manual4.3.0.72User Manual 4.3.0.71 is installed locally and is fully applicable to NOTOCORD-hem version 4.3.0.72.Documentation2015/11/03
NDSI70ab4.3.0.71Support has been added for implant L00.DSI70ab2015/07/30
NTMS31a4.3.0.71The server supports SaO2 channels of PORTI devices.TMS31a2015/07/30
NInstaller4.3.0.71Installs Microsoft .NET Framework 4.5 if not found on target machine, required by DSI71a.DSI71a2015/07/30
NDSI71a/b/c/d/e4.3.0.71DSI Link2 acquisition server acquires up to 4/8/12/16/32 implants.DSI71a2015/07/30
NVideo4.3.0.71Video synchronization works also for DSI70a and DSI71a.KRN43a2015/07/30
NMarkers4.3.0.71Rendering speed for displaying large amount of event markers is highly improved.KRN43a2015/07/30
NVOL11a4.3.0.71The module calculates the geometric mean of three inputs.VOL11a2015/07/30
CTMS31a4.3.0.71Acquisition starts with a configuration containing the maximum number of output channels allowed by the license with this correctionWI49642015/07/30
CDSI60a/b/c/d4.3.0.71Support for HD implants has been added for DSI OpenART 4.34.WI48812015/07/30
NTMS31a4.3.0.71The server supports REFA HR with a sampling rate until 5 kHz.TMS31a2015/07/30
NDSI70ab4.3.0.71During data collection, when APR device is not responding, the last measured value is applied instead of the default value.WI50182015/07/30
NTMS31a4.3.0.71The server supports Windows 7 64-bit driver.TMS31a2015/07/30
CDSI70ab4.3.0.71Acquisition starts when multiple instances are configured to have a delayed start at the same time point with this correction.WI51492015/07/30
CDSI70ab4.3.0.71When an instance is configured to have a delayed start while another instance is in acquisition, this correction prevents any memory overflow of the server.  WI51382015/07/30
CDSI70ab4.3.0.71Acquisition starts in scheduled recording mode with this correction.WI48602015/07/30
CDSI70ab4.3.0.71During data collection, the ambient pressure value captured by the APR device is used to correct the pressure channels instead of the default ambient pressure value with this correction.WI49682015/07/30
NDSI60a/b/c/d4.3.0.70Support for DSI OpenART 4.34 on 32-bit and 64-bit systems has been added.DSI60a/b/c/d2015/02/18
NLicence manager4.3.0.70There is now support for deployment licenses of custom EXT10anm modules created by customers.KRN43a2015/02/18
CCTD60v4.3.0.70With this correction, multiple video exports can be performed from CTD50v (HikFormat) to completion without stopping.WI24182015/02/18
CDSI60a/b/c/d4.3.0.70The server interface is now in English instead of French with Region and Language format set to English UK or US.WI44882015/02/18
CDSI60a/b/c/d4.3.0.70Permissions have been corrected on %ProgramData%\NOTOCORD so that users other than Administrator and the first user to use DSI60 can open a DSI60 configuration. This corrects the same problem for DSI70.WI48512015/02/18
CDSI70ab4.3.0.70With this correction, the server no longer crashes when the Ambient Pressure Reference (E2S-1/APR-1) is connected to the Physiotel network.WI46602015/02/18
CExcel wizard4.3.0.70With this correction, when Office 2013 is installed, starting NOTOCORD-hem Evolution 4.3 installs the Excel wizard add-in into Excel 2013.WI44732015/02/18
CInstaller4.3.0.70Shortcut, directory, and file names no loger contain the trademark (TM) symbol which caused problems on systems using the unsupported Japanese Region and Language settings.WI44702015/02/18
CIPD30a4.3.0.70When an animal ID is changed, it is now correctly updated throughout the user interface with this correction.WI44492015/02/18
CIPD30a4.3.0.70With this correction, the previous log is no longer incorrectly restored after modifying the sequence template and then clicking on Ok in the properties window.WI44692015/02/18
KDRV10a4.3.0.70Recalculate only DRV10a, then the module is in acquisition mode (green), and its properties cannot be changed. The workarounds are to recalculate an upstream module, recalculate everything, or close and open the NSS file.WI24082015/02/18
KECG51a4.3.0.70Increasing the QT Duration Max can produce an access violation error message when retrying to propagate marks. The workaround is to lower the QT Duration Max to, for example, 500ms, recalclate all, close and then reopen the NSS file.WI12032015/02/18
NExcel4.3.0.69The Excel wizard works under Excel 2013 with Microsoft hot fix KB2863878. This hot fix is integrated into the installer.WI32932014/09/15
NEEG10b4.3.0.69Spectral analysis of electroencephalogram.EEG10b2014/09/15
NIPD30ab4.3.0.69iPRECIO® Dual pump acquisition server and controller.IPD30ab2014/09/15
NRME10a4.3.0.69Reference mark editor for arrhythmia now with GLP features.RME10a2014/09/15
NSTE10a4.3.0.69Stellar Telemetry acquisition module.STE10a2014/09/15
NSZR30a4.3.0.69Automated seizure detector.SZR30a2014/09/15
NTHR10c4.3.0.69Threshold detection with e-mail alarm.THR10c2014/09/15
NVMA10a4.3.0.69Analysis of respiratory parameters in freely moving animals now excepts input from CVT10a (Signal Format Converter).VMA10a2014/09/15
NWPI10a4.3.0.69World Precision Instruments acquisition module.WPI10a2014/09/15
RCorscience_BT3-64.3.0.69Bluetooth® External Telemetry BT3/6 ECG device from Corscience.Corscience_BT3-62014/09/15
RCSW10ab4.3.0.69Corscience BT3/6 acquisition up to 7/14 devices.CSW1ab2014/09/15
RINS30a4.3.0.69Customer specific Instrunet acquisition server.INS30a2014/09/15
RSVR10a4.3.0.69Systemic vascular resistance calculation.SVR10a2014/09/15
RVIV10a4.3.0.69Vivometrics LifeShirt Acquisition server.VIV10a2014/09/15
RMisc4.3.0.69Miscellaneous removal of very old modules: RAW30a, RCD30d, SWT10.Misc2014/09/15
CConcatenate4.3.0.69After opening an NSS file, opening concatentate, selecting a destination file, and then choosing cancel, the destination file is no longer created nor remembered with this correction.WI22742014/09/15
CConfiguration report4.3.0.69For a newly created NSS file followed by loading a configuration, the configuration report correctly displays the name of the configuration file with this correction.WI12002014/09/15
CCTD60a4.3.0.69One can now copy the X and/or Y axis of multiple graphs and paste to another application, e.g,. paint, with this correction.WI20282014/09/15
CDSI60a/b/c/d4.3.0.69An acquisition no longer fails if an animal ID is longer then 7 characters. Only the first seven characters are used as before so they still must represent a unique ID with this correction.WI21692014/09/15
CDSI70ab4.3.0.69With this corrections, 50/60 Hz filtering occurs in the neighborhood of QRS complexes.WI27052014/09/15
CDSI70ab4.3.0.69With this correction, the installer now installs the required Visual C++ runtime distributable prerequisites if not present that allow the ambient pressure device to be detected.WI30952014/09/15
CDSI70ab4.3.0.69When "Save Raw Data" is selected and "Save filtered data" is unselected, all data streams, raw and filtered, are no longer removed with this correction.WI31432014/09/15
CDSI70ab4.3.0.69With this correction, when the pressure decreases under the lower limit of the domain, the correct value is now displayed.WI31552014/09/15
CECG30s4.3.0.69The "Time period of QT stablization" field can now display clearly three digits with this correction.WI12512014/09/15
CECG31p4.3.0.69"Invalid floating point operation" for certain signals and for long acquistions with crashes of the software from time to time no longer occur with this correction.WI42382014/09/15
CECG51a4.3.0.69When a mark is renamed from "X" to "Y", the associated "Missing - X" is renamed to "Missing - Y" with this correction.WI12062014/09/15
CECG51a4.3.0.69Statistics are displayed immediately after using "Apply marks from all marked reference beats" instead of after the next manual mark edition with this correction.WI12082014/09/15
CLicence manager4.3.0.69The software cannot start if the licence manager is not running. An error message to this effect is displayed when it is not running with this correction instead of displaying the message that a license is not present.WI12672014/09/15
CLVP31a4.3.0.69With eight instances of Hem.exe running with eight LVP31a modules per instance, an entry level system can acquire up to seven days of data with this correction.WI12072014/09/15
CQTC10a4.3.0.69QTC10a no longer produces error codes E0010003 nor E0010004 in the log file when ECG51a is present in a configuration during acquisition.WI12052014/09/15
CRIA30a4.3.0.69With this correction, the calibration factor is updated after a reculculation for a zone starting at anytime, not just near the start time.WI12092014/09/15
CSecurity and audit trail4.3.0.69The correct reviewed or unreviewed status is indicated when an NSS file is read-only with this correction.WI11992014/09/15
CSecurity and audit trail4.3.0.69Licens numbers, e.g., 8768-900536, are no longer repeated multiple times in the audit trail.WI22312014/09/15
KDTV10a4.3.0.69This module is sensitive to the manner in which data is fed into it and for certain cases can yield an exception in the log file instead of a result.WI11982014/09/15
KHRG30a4.3.0.69For certain signals, incorrect detection of successive cycles is observed with this module. Contact support@notocord.com for a workaround using OPR10a.WI19782014/09/15
KTCF30a4.3.0.69If the input signal is inverted, the cardiac output is not calculated and the beginning and manximum triggers are incorrect. Contact support@notocord.com for a workaround.WI12502014/09/15
KDSI60a/b/c/d4.3.0.69This module does not support OpenART's inventory feature. Please use the recycle bin. Contact support@notocord.com for more information.TD21592014/09/15
CDSI70a/b4.3.0.67An acquisition is no longer limited to 64 channels. It is now limited to 256 channels.WI19032013/01/23
NDSI70a/b4.3.0.66DSI PhysioTel™ Digital acquisition server up to 4/8 devicesDSI70ab2013/01/21
NISO10r4.3.0.66Rabbit isolated heart electrocardiogram analyzerISO10r2013/01/21
NLVP31a4.3.0.66Left ventricular pressure analyzerLVP31a2013/01/21
NTRI10a4.3.0.66Analysis of transmural repolarization indexTRI10a2013/01/21
CRIA30a4.3.0.66Corrected the update of the calibration factor that did not update for zones late in the acquistion when recalculating.WI12092013/01/21
NARR30a4.3.0.47Arrhythmia detectorARR30a2012/05/30
NBRS10A4.3.0.47Baroreflex sequence analyzerBRS10A2012/05/30
NCSA10a4.3.0.47Cross-spectrum analyzerCSA10a2012/05/30
NRME10a4.3.0.47Reference mark editor for arrhythmiaRME10a2012/05/30
RPRT30a4.3.0.47The continuous thermal printer AR200 signal recorder module was removed.PRT30a2012/05/30
CAbout
window
4.3.0.47Eliminated a repetition of the license number in the About window. Use the TokenManager to see complete license details.TD27112012/05/30
CInstaller4.3.0.47Corrected the installer so that permissions on files in the workfile directory do not disappear after a GLP upgrade.TD27172012/05/30
CPLY30h4.3.0.47Corrected an incorrect average value of the Tidal volume.TD27312012/05/30
CSTT30a4.3.0.47Corrected a spelling error in the user interface that caused confusion.TD27002012/05/30
CVME10e/v4.3.0.47Corrected for the multi-graph window not repainting when the configuration window is open. This error generates log file entries and can block the user interface.TD27592012/05/30
CVSH10a/b4.3.0.47Corrected an error in reading the license that prevented it from launching the VSH server.TD27152012/05/30
CXYD30a4.3.0.47Reference document corrected and udpated concerning the display of successive zones with differing frequencies.TD24222012/05/30
KAMG50a4.3.0.47Under rare circumstances, an audit trail entry is created while attempting a disallowed operation.TD16742012/05/30
KAMG50a4.3.0.47Domain users and groups within a local group are not recognised even though the local group is authorised to perform an action.TD20612012/05/30
KAMG50a4.3.0.47When a workstation no longer has network access to the access control server, it uses the latest access control configuration saved locally.TD20982012/05/30
KCTD60a4.3.0.47The slope in the results window is calculated between the lower left corner and the upper right corner of the zone and not between the first time point and the last time point of the signal in the zone.TD22582012/05/30
KCTD60a4.3.0.47With two screens, the Configuration Setup window and CTD60a dialog box do not appear on the same screen.TD26302012/05/30
KCTD60a4.3.0.47When CTD60a is configured with two columns, one graph per column, if one graph is deleted, it is not possible to add a new graph.TD26742012/05/30
KECG51a4.3.0.47Recommendation: Do not use two-column mode with ECG51a. In two-column mode, when deleting the last graph with the right mouse-click menu, ECG51a and NOTOCORD-hem Evolution terminate. Reference beats may be lost and modules may disconnect from ECG51a.TD27482012/05/30
KKRN43a4.3.0.47When a module and its license are added to a file after acquisition, the module is not available under a review license.TD25642012/05/30
KKRN43a4.3.0.47A configuration created on a system with two screens may not be visible on a system with one screen.TD26102012/05/30
KLMG10a4.3.0.47When newly created, local groups cannot be immediately used for video access control, a reboot is required.TD18182012/05/30
KNVC31a4.3.0.47Parameters for this module are not written to the audit trail or to a configuration report.TD21702012/05/30
KVSB10a4.3.0.47The error message "Video Server does not respond" appears when there are two network cards present and one of them is connected to the network of the Windows domain.TD22782012/05/30
NUser manual4.3.0.20The user manual and system requirements are updated for Windows 7.Documentation2011/11/25
NWindows 7 32/64-bit4.3.0.20NOTOCORD-hem Evolution 4.3 supports Windows 7 32/64-bit systems.W7-compatibility2011/11/25
NOpenART 4.324.3.0.20NOTOCORD-hem Evolution 4.3 supports HD-S21 implants and requires OpenART 4.3.DSI2011/11/25
CVSH10a/b4.3.0.20The export video function on a zone did not work with the latest Hikvision drivers and SDK.TD23512011/11/25
RDiskette licenses4.3.0.20NOTOCORD-hem Evolution 4.3 no longer uses diskette licensing.TokenManager2011/11/25